Sustitución del Ni por Mn en aceros usados en implantes quirúrgicos
Abstract
The Ni present in 316L stainless steel is considered as a possible cause of allergenic reactions in the human organism. For this reason, the present work evaluates the possibility of replacing Ni by Mn, which also is an austenite stabiliser.The chemical composition of the analysed steel was as follows: 0,04%C, 17,9%Cr, 21%Mn, 1,8%Mo. As cast microstructure consisted in an austenitic matrix with delta ferrite and sigma phase in the interdendritic regions. The material was submitted to a solution annealing, from which the mechanical properties and the susceptibility to localised corrosion were evaluated. The tensile strenth was 706 MPa, higher than AISI 316L. On the other hand, the electrochemical test indicated that Mn steel showed a higher tendency to pitting, because the presence of delta ferrite in the microstructure.
